68 CFS/CCS-BA-e-1220
14 Configuration bulk material balance / technical parameters
Changes should only be carried out by trained specialized personnel.
Navigation in menu see chap. 13.1
In factory the balance KERN CFS or the counting system KERN CCS are
preconfigured in a way that normally no more changes will be necessary.
But if there are special conditions of use or if as bulk material balance an other
weighing bridge (not preconfigured by KERN) is connected, in the menu block
" " the required settings can be made.
Technical data
Supply voltage:
5 VDC
Max. signal voltage
0-20 mV
Zeroing range
0-5 mv
Sensitivity
> 0.02 μv
Resistance parameter
87 O Min., 4 x 350 Ω load cell
Connection
4 poles
Max. cable length
6 meter
Connection plug
9 pin d-subminiature bushing
Vedere la pagina 67